PRODUCT INFORMATION
Online Conductivity / Resistivity Meter
CON5101-TQdegassed cation conductivity online analyzer
Xuetonginstrument
Pre-sales and after-sales support
technical parameters
- 1, instrument class: 1.0;
- 2, conductivitymeasurement:
- 0.01Cell constant: (0.055–0.200)μS/cm, Resolution: 0.001μS/cm
- (0.055–2.000)μS/cm, Resolution: 0.001μS/cm
- (0.055–20.00)μS/cm, Resolution: 0.01μS/cm
- 0.1: Cell constant: (0.055–200.0)μS/cm, Resolution: 0.1μS/cm
- 3, measurementsystem: three (K=0.01) conductivity electrode, built-in PT1000 temperatureelectrode;
- 4, temperaturemeasurement: (0.0– 100)℃, Resolution: 0.1℃;
- 5, temperaturecompensation mode: linear(ordinary water), neutral salt, ammonia, strong acid, compensationtemperature(0–99.9)℃;
- 6, pH calculation range: (7.5– 11.5)pH;
- 7, pH calculate samplerequirements:
- only 1 alkalinereagent
- main pollutants come from NaCL
- phosphate<0.5mg/L
- if pH<8, pollutantsconcentrationalkaline reagent consumption must be very low
- 8, concentrationcalculation range: 0.01-10mg/L;
- 9, electronic uniterror:
- indication: conductivity 0.25%FS
- temperaturecompensation: 0.25×10-2%/10℃
- Repeatability: 0.25%FS
- Stability: 0.25%FS/24h
- response time: ≤30s
- 10, complete instrumentperformanceerror:
- indication: 1.5%FS
- temperature: 0.5℃(0–60)℃
- repeat: 0.7%FS
- 11, currentoutput:
- (0.00–20.00) mA
- (0.00– 10.00) mA
- (4.00–20.00)mA three options, dual-channel 800Ωload, error 0.1mA
- 12, data storage: ≥10 0,000 points;
- 13, sample conditions:
- flow: (9– 15)L/h
- pressure: ≤0.2MPa
- sampleoperating solids content: ≤5μm(no colloids permitted, oil/grease)
- 14, environmental conditions: temperature:(5–40)℃, Relative humidity: ≤95%;
- 15, Power supplyrequirements: AC(100–240)V, (50–60)Hz, 3500W;
- 16, appearancedimensions: 520mm*1020mm;
- 17, Mounting: wall-mount;
- 18, Weight: ≤20kg;
- 19, safetyrequirements:
- leakagecurrent: ≤3.5mA
- earthresistance: ≤0.1Ω
- insulation resistance: electronic unit Power supply, line-to-earth insulationresistance≥40MΩ
- dielectric strength: Power supply, between line and earth, Power supplybetween the input terminals and accessible metal parts, shall withstand AC 1500V, 50Hz testvoltage, duration 1min, there shall be no breakdown or flashover
- ESD test: panel, enclosure, cable test in air 8kV, contact 4kV, duration 1min, no lock-up