PRODUCT INFORMATION
Online Conductivity / Resistivity Meter
CON5102 intelligent online conductivity meter
Xuetonginstrument
Pre-sales and after-sales support
technical parameters
- Standard: JB/T 6855-2017《industrialconductivity meter》;
- conductivity Measuring range: 0.0–20μS/cm(with 0.01electrode); 0.1–200μS/cm(with 0.1electrode); 1.0–2000μS/cm(with 1.0 electrode); 10–20000μS/cm(with 10.0electrode);
- Accuracy: conductivity: ±1.0%FS; temperature ±0.5℃;
- Electronic unit basic error: conductivity±0.5%FS, temperature ±0.3℃(0–60℃);
- Automatic temperature compensation range: 0–99.9℃, referenced to 25°Ctemperature;
- Sample water: 0–99.9℃, 0.6MPa;
- Instrument basic error: conductivity ±1.0% FS, temperature ±0.5℃(0–60℃);
- electronic unit automatictemperaturecompensation error: ± 0.5%FS;
- Electronic unit repeatability error: ±0.2%FS±1 digits;
- electronic unit Stability: ±0.2%±1 digitsFS/24h;
- electronic unitalarmerror: ≤±3.0% FS;
- electronic unit Ambient temperatureinfluence error: ≤±0.5% FS;
- electronic unit Power supplyvoltage influence error: ≤±0.3% FS;
- current Isolated output: 4–20 mA(load<750 Ω );
- electronic unit Output current error: ≤±1.0% FS;
- alarm relay: AC220V, 3A;
- RS485communication interface(optional);
- Power supply: AC220V±22V, 50Hz±1Hz; optionalDC24±2.4V,1Apower supply; please specify when ordering;
- Ingress protection: IP65;
- Overall dimensions: 159 (L)×159 (W)×138 (D)mm, Cutout size: 138×138 mm;
- Weight: 1.0kg;
- Operating conditions: Ambient temperature0–60℃, Relative humidity<85%.