PRODUCT INFORMATION
Online Conductivity / Resistivity Meter
CON5101-GY online conductivity meter (inductive)
Xuetonginstrument
Pre-sales and after-sales support
Technical specifications
- Standard: JB/T 6855-2017《industrialconductivity meter》;
- conductivity Measuring range: (0–2.00)mS/cm; (0–20.00)mS/cm; (0–200.0)mS/cm; (0–500.0)mS/cm; (0–1000)mS/cm; (0–2000)mS/cm;
- Electronic unit basic error: conductivity: ±1.0%FS, temperature: ±0.3℃(0–60℃);
- Automatic temperature compensation range: 0–99.9℃, referenced to 25°Ctemperature;
- Sample water: 0–99.9℃, 0.6 MPa;
- Instrument basic error: ±2.0%FS, temperature: ±0.5℃(0–60℃);
- electronic unit automatictemperaturecompensation error: ±0.5%FS;
- Electronic unit repeatability error: ± 0.2%FS±1 digits;
- electronic unit Stability: ±0.2%FS±1 digits/24h;
- current Isolated output: 0–10mA(load<1.5kΩ ), 4–20 mA(load<750Ω );
- Output current error: ≤±1%FS;
- electronic unit Ambient temperatureinfluence error: ≤±0.5%FS;
- electronic unit Power supplyvoltage influence error: ≤±0.3%FS;
- alarm relay: AC220V, 3A;
- RS485communication interface(optional);
- Power supply: AC220V±22V, 50Hz±1Hz; optionalDC24±2.4V,1Apower supply;
- Ingress protection: IP65;
- clock Accuracy: ± 1 min/month;
- data storage capacity: 1 months(1 pt/5 min);
- data retention after power loss: 10 years;
- Overall dimensions: 159 (L)×159 (W)×138 (D)mm; Cutout size: 138×138 mm;
- Weight: 0.8kg;
- Operating conditions: Ambient temperature0–60℃ Relative humidity<85%;
- with inductiveconductivity electrode.