Professional water quality analyzers for environmental and industrial monitoring
gyxtyb@126.com +86 851 8799 0163 CN
PRODUCT INFORMATION
Online Conductivity / Resistivity Meter

CON5101-TDS online TDS meter

Xuetonginstrument Pre-sales and after-sales support

Technical specifications

  • Standard: JB/T 6855-2017《industrialconductivity meter》;
  • conductivity Measuring range: (0.0–10)mg/l; (with 0.01electrode); (0.0–100)mg/l; (with 0.1electrode); (0.0–1000)mg/l; (with 1.0 electrode); (0.0–10000)mg/l; (with 10.0electrode);  
  • Electronic unit basic error: TDS ±1.0%FS, temperature: ±0.3℃(0–60℃);  
  • Automatic temperature compensation range: 0–99.9℃, referenced to 25°Ctemperature;
  • Sample water: 0–99.9℃, 0.6MPa;
  • Instrument basic error: TDS ±1.0%FS; temperature: ±0.3℃(0–60℃);
  • electronic unit automatictemperaturecompensation error: ±0.5%FS;  
  • Electronic unit repeatability error: ± 0.2%FS±1 digits;
  • electronic unit Stability: ±0.2%FS±1 digits/24h;
  • current Isolated output: 0–10mA(load<1.5kΩ ), 4–20 mA(load<750Ω );
  • Output current error: ≤±1%FS;
  • electronic unit Ambient temperatureinfluence error: ≤±0.5%FS;
  • electronic unit Power supplyvoltage influence error: ≤±0.3%FS;
  • alarm relay: AC220V, 3A;
  • RS485communication interface(optional);
  • Power supply: AC220V±22V, 50Hz±1Hz; optionalDC24±2.4V,1Apower supply;
  • Ingress protection: IP65;
  • clock Accuracy: ± 1 min/month;
  • data storage capacity: 1 months(1 pt/5 min);
  • data retention after power loss: 10 years;
  • Overall dimensions: 159 (L)×159 (W)×138 (D)mm; Cutout size: 138×138 mm;
  • Weight: 0.8kg;
  • Operating conditions: Ambient temperature0–60℃   Relative humidity<85%;
  • electrode-typeconductivity electrode.