PRODUCT INFORMATION
Online Conductivity / Resistivity Meter
CON5103B online conductivity meter
Xuetonginstrument
Pre-sales and after-sales support
Technical specifications
- Standard: JB/T 6855-2017《industrialconductivity meter》;
- LCD, operator interface;
- conductivity Measuring range: 0.01–30μS/cm(with 0.01electrode); 0.1–300μS/cm(with 0.1electrode); 1.0–3000μS/cm(with 1.0 electrode); 10–30000μS/cm(with 10.0electrode);
remark: wide rangeinstrumentabove Measuring range expanded from 10 times, with 30electrodemeasurement600ms/cm, please specify when ordering. - Electronic unit basic error: conductivity: ±0.5%FS, temperature: ±0.3℃(0–60℃);
- Automatic temperature compensation range: 0–99.9℃, referenced to 25°Ctemperature;
- Sample water: 0–99.9℃, 0.6MPa;
- Instrument basic error: conductivity: ±1.0%FS, temperature: ±0.5℃(0–60℃);
- electronic unit automatictemperaturecompensation error: ±0.5%FS;
- Electronic unit repeatability error: ± 0.2%FS±1 digits;
- electronic unit Stability: ±0.2%FS±1 digits/24h;
- current Isolated output: 4–20 mA(load<750Ω );
- Output current error: ≤±1%FS;
- electronic unit Ambient temperatureinfluence error: ≤±0.5%FS;
- electronic unit Power supplyvoltage influence error: ≤±0.3%FS;
- alarm relay: AC220V, 7A;
- RS485communication interface(optional);
- Power supply: AC220V±22V, 50Hz±1Hz; optionalDC24±2.4V, 1Apower supply;
- Ingress protection: IP65;
- data retention after power loss:10 years;
- Overall dimensions: 96 (L)×96 (W)×110 (D)mm;
Cutout size: 92×92mm; - Weight: 0.8kg;
- Operating conditions: Ambient temperature0–60℃ Relative humidity<85%;
- optional 0.01, 0.1, 1.0, 10.0four typesconductivity electrode.