Infrared Oil Analyzer
infrared oil analyzer
Xuetonginstrument
Pre-sales and after-sales support
Technical specifications
- 1.Method detection limit When the sample volume is 500 ml and the extract volume is 50 ml, the sample detection limit is 0.04 mg/L
- 2.Instrument detection limit DL≤0.03mg/L(3×SD from 11 blank measurements)
- 3.measurementaccuracy ≤±2%
- 4.instrument Repeatability RSD≤0.5%(30-60mg/Loil standard measurement11); RSD≤5%(0.2mg/Lsample)
- 5.basic Measuring range 0.0–800.0mg/L
- 6.lowest detectionconcentration 0.004mg/L(enrichment measurement)
- 7.maximummeasurementconcentration 800000mg/L(dilution measurement)
- 8.absorbancerange 0.0000–2.0000AU(i.e. transmittance100–1%T)
- 9.scanwavenumberrange 3400cm-1–2400cm-1(2941nm–4167nm)
- 10.measurementwavelength range 3200nm–3600nm
- 11.wavenumberaccuracy ±1cm-1
- 12.wavelength Repeatability ±1nm
- 13.linearrelatedcoefficient R>0.999
- 14.scan speed full-spectrum scan3200nm–3600nm, 30 s to completeone samplemeasurement
- 15.main unit Weight 12Kg
- 16.operating environment -5–45℃, Relative humidity: 20–80%
- 17.main-unit outlinedimensions 450mm (L)×320mm (W)×130mm (H)
- 18.main unit Power supplypower 220 ± 10% V, 50Hz, 50W