PRODUCT INFORMATION
Online Dissolved Oxygen Meter
OXY5401-IXDDL dual-channel dissolved oxygen meter
Xuetonginstrument
Pre-sales and after-sales support
Technical specifications
- 1, Standard: SC/T 7006-2001《dissolved oxygen analyzer》;
- 2, Measuring range: (0–200.0)ug/L; (0–20.00)mg/L(auto-ranging); (0–60)℃; saturation: (0–200.0)%;
- 3, resolution: 0.1 ug/L, 0.01 mg/L, 0.1℃; 0.1%;
- 4, complete instrument Basic error: ug/L: ±1.0%FS; mg/L: ±0.5%FS, temperature: ±0.5℃;
- 5, complete-instrument indication Repeatability: ±0.5%FS;
- 6, complete-instrument indication Stability: ±1.0%FS;
- 7, Automatic temperature compensation range: (0–60)℃, referenced to 25°C;
- 8, response time: <60 s( of full scale98%, 25℃)37℃: 98%full scale<20 s;
- 9, Display: on the same screendisplaytwo channel dissolved oxygen;
- 10, Output current error: ≤±1.0%FS;
- 11, Isolated output: (0–10) mA(loadresistance<1.5KΩ ); (4–20)mA(loadresistance<750Ω );
- 12, RS485communication interface(optional);
- 13, data retention after power loss: 10 years;
- 14, alarm relay: AC220V, 3A;
- 15, Power supply: 220V±10% 50±1HZ; optionalDC24V, please specify when ordering.
- 16, Overall dimensions: transmitter: 159 (L)*159 (W)*138 (D)*mm; Cutout size: (138*138)mm
- 17, Weight: transmitter: 0.8Kg;
- 18, Operating conditions: Ambient temperature: 0–60℃, Relative humidity: <85%;
- 19, inlet/outlet piping: stainless steelrigid-pipe connection; optionalφ8stainless steelquick connector.
- 20, flowrequirements: (100–400)ml/min(6–24)L/h.