PRODUCT INFORMATION
Online Sodium Analyzer
DWG5301C online sodium analyzer (with pH)
Xuetonginstrument
Pre-sales and after-sales support
Technical specifications
- 1.Standard: JB/T 6858-2018 Test methods for pH meters and ion meters;
- 2.Measuring range: 0–100ug/L, 0–23000mg/L, 0.00–14.00pH;
- 3.Resolution: 0.01 µg/L, 0.01mg/L; 0.00p Na–9.00p Na, Resolution: 0.01p Na; 0–60℃, Resolution: 0.1 ℃; pH: 0.00–14.00, Resolution: 0.01;
- 4.Basic error: ± 2.5%, temperature±0.3℃;
- 5.Automatic temperature compensation range: 0–99.9℃, referenced to 25°C;
- 6.Electronic unit temperature compensation error: ±2.5%;
- 7.Electronic unit repeatability error: of reading ±2.5%;
- 8.Stability: of reading ±2.5%/24h;
- 9.Input impedance: ≥1012 Ω;
- 10.Sample water: 0–60℃, 0.6Mpa, Sample flow: 30ml/min–300ml/min;
- 11.Isolated output: 4–20mA(load<750Ω );
- 12.Output current error: ≤±1%FS;
- 13.Alarm N/O contact: AC250V, 7A;
- 14.Power supply: 220V±10%, 50±1Hz;
- 15.Standard enclosure dimensions:
- Panel-cutout mounting: Cutout size: 712mm (H)*392mm (W), Panel locating-hole size: 739mm (H)*300mm (W), Mounting hole: Φ12mm, Overall dimensions: 770mm (H)*440mm (W).
- Wall mounting: Hole-center distance: 588mm(294mm+294mm) (H)*414mm (W), Mounting hole: Φ12mm, Overall dimensions: 770mm (H)*440mm (W).
- 16.Large enclosure dimensions:
- Panel-cutout mounting: Cutout size: 836mm (H)*436mm (W), Panel locating-hole size: 470mm (H)*456mm (W), Mounting hole: Φ8mm, Overall dimensions: 890mm (H)*485mm (W).
- Locating-hole size: 739 (H)*300 (W), Four-hole pattern: hole diameterФ12; Wall-mount overall dimensions: 750 (H)*380 (W)*234 (D); Locating-hole size: 730 (H)*280 (W);
- 17.Four-hole pattern: hole diameterФ12;
- 18.Weight: 25kg;
- 19.Operating conditions: Ambient temperature: 0–60℃, Relative humidity: <85%.
- 20.Reagent consumption: Typical diisopropylamine consumption is 1000 mL/month; actual use depends on operating hours.
Functions
- 1, Includes pH measurement for the actual sample pH.
- 2, When pH falls below 12.0 the booster module air pump starts automatically to maintain sample alkalization.
- 3, The pH control band is configurable so the air-pump start pH can match site requirements.
- 4, A float switch keeps the air pump off when no sample is present, reducing diisopropylamine use; measurement starts automatically when sample returns.
- 5, Constant-flow design for more stable readings.