PRODUCT INFORMATION
Online Calculated pH Meter
pHG5201B online calculated pH meter
Xuetonginstrument
Pre-sales and after-sales support
Overview
pHG5201Bonline calculated pH meter, Calculates pH from conductivity according to the German VGB guideline relating specific and cation conductivity to pH.pHG5201Bonline calculated pH meter Suitable for thermal power plant feedwater (economizer) and condensate water plus ammonia pHonlinecontinuous monitoring.
pHG5201Bonline calculated pH meter, via measurementconductivitycalculates pH, eliminates replacement of pH electrode and pH electroderequired monthlyperform calibration, this both saves pH electrode cost, and also reduces labour cost, further increases measurement accuracy and Stability.
Technical specifications
- pHMeasuring range: (7.50–10.50)pH;
- LCD, operator interface;
- conductivity Measuring range: 0.01–20μS/cm(with 0.01electrode), typically achievable100.0μS/cm; 0.1–200μS/cm(with 0.1electrode), typically achievable1000.0μS/c;
- Electronic unit basic error: conductivity: ±0.5%FS, temperature: ±0.3℃(0–60℃);
- Automatic temperature compensation range: 0–99.9℃, referenced to 25°Ctemperature;
- Sample water: 0–99.9℃, 0.6MPa;
- Instrument basic error: conductivity: ±1.0%FS, temperature: ±0.5℃(0–60℃);
- electronic unit automatictemperaturecompensation error: ±0.5%FS;
- Electronic unit repeatability error: ± 0.2%FS±1 digits;
- electronic unit Stability: ±0.2%FS±1 digits/24h;
- current Isolated output: 4–20 mA(load<750Ω );
- Output current error: ≤±1%FS;
- electronic unit Ambient temperatureinfluence error: ≤±0.5%FS;
- electronic unit Power supplyvoltage influence error: ≤±0.3%FS;
- alarm relay: AC220V, 7A;
- RS485communication interface(optional);
- Power supply: AC220V±22V, 50Hz±1Hz; optionalDC24±2.4V,1Apower supply;
- Ingress protection: IP65;
- data retention after power loss:10 years;
- Overall dimensions: 96 (L)×96 (W)×110 (D)mm; Cutout size: 92×92mm;
- Weight: 0.8kg;
- Operating conditions: Ambient temperature0–60℃ Relative humidity<85%.
Standard configuration
| pHG5201Btransmitter | one |
| measurement | one |
| XT-DDJ-0.01conductivity electrode | one |
Features
- fast sensor response;
- sensor measurement Stability;
- sensorlow maintenance;
- instrumentmeasurementaccurate data, Stability;
- microprocessor-based, uses high-performanceCPUchip, AccuracyADconversion technology and SMTSMT technology;
- EMC(EMC/RFI)design: designed to European StandardEN50081/50082design and manufacture;
- dustproof and waterproofdesign;
- calculatedpH metercannot uses pHStandardperform calibration, instrumententer or correct in parameter setup Cell constantthen.