Professional water quality analyzers for environmental and industrial monitoring
gyxtyb@126.com +86 851 8799 0163 CN
PRODUCT INFORMATION
Online COD, NH3-N, TP & TN

XT-IV total nitrogen online automatic water quality analyzer

Xuetonginstrument Pre-sales and after-sales support

Technical specifications

  • Method: sulfuric acidpersulfate oxidation UV spectrophotometry; water quality total nitrogen determination alkaline sulfuric acidpotassium persulfate digestion UV spectrophotometry (HJ 636—2012)《total nitrogenautomatic water quality analyzertechnical requirements》(HJ/T 102-2003), 《water qualitytotal nitrogenonlineanalyzer operating instruction(reference《water quality total nitrogenonlineanalyzer requirements and test method》)》(CNEMC-03-ZJZX-005-2023).
  • Measuring range:  0 mg/L–20mg/L, 0 mg/L–50mg/L, 0 mg/L–100 mg/L, 0 mg/L–200 mg/L, 0 mg/L–500mg/L(rangeexpandable).
  • Repeatability: ≤ ±2%.
  • 24h 20%rangeupper limit: 0.05mg/L; 24h 80%rangeupper limit± 1%.
  • indicationerror ± 10 %.
  • linearity: ≤ ±10%.
  • limit of quantitation: ≤ 0.30 mg/L.
  • memory effect: standard solution80%→20%≤±0.40mg/L; standard solution20%→80%≤±0.80mg/L.
  • voltage influence: ≤ ±5%.
  • Ambient temperatureinfluence: ≤ ±5%.
  • real-sample comparison: real sampletotal nitrogen<2 mg/L (concentration1.5 mg/L Standardsamplereplacereal sampleperform test)± 0.2 mg/L; real sampletotal nitrogen≥2 mg/L, ≤ ± 10 %.
  • data validity: ≥90%.
  • consistency: ≥90%.
  • waste-liquid separation Functions.
  • standard-solution check Functions.
  • spike recovery Functions.
  • fault reset and power restoration, the analyzer automatically discharges residual reagents, automatic cleaning, automatically resets to restart the test Functions.
  • alarms and power loss do not cause data loss.
  • outputretained sample, triggered sampling Functions.
  • reagent-low alarm, component-fault alarm, sample-fault, over-limit and other alarms Functions.
  • touchscreendisplay and command input.
  • sampling cycle: time interval(10 min–9999 minfreely adjustable) and on the hourmeasurementmode.
  • calibration interval: 1–99adjustable to any interval and time of day, automatic calibration, automatic calibration.
  • data storage: savedata storagemore than 10 years, historical records can be recalled at any time.
  • output: 1-channel national-standard 212 RS-232, 1-channel Mod-Bus RS-232(RS-485), 1-channel 4 –20 mA.
  • Power supply: AC220±10% V, 50±10% Hz, 5 A.
  • dimensions: 1500 mm×500 mm×450mm(×× D).
  • environmentrequirements: Ambient temperature: 5℃–40℃; temperatureadjustable indoor, recommendedtemperature25 ℃–28 ℃; humidity≤90% RH(non-condensing).
LET’S TALK WATER