OXY5401-IXDDL dual-channel dissolved oxygen meter
Overview
OXY5401-IXDDLdual-channel dissolved oxygen meteris manufactured by Xuetong onlineelectrochemicalanalyzer, highly intelligent onlinecontinuous trace-oxygen monitor. can be fitted with 2XT-DOPPb-DDLcurrentelectrode, measurementtrace oxygen in two sample channels, automaticfrom ppbppmwide rangemeasurement, is used to monitor boiler feedwater, condensate, deaeratoroutletwater quality oxygen content in measurement dedicated instrument.
XT-DOPPb-DDLcurrentelectrodefast response, stable, reliable, and low operating cost Features, suitable for large-scale use in thermal power plants. instrumentwith full, Functions, so that it “practical, convenient, Functionscomplete” and other advantages in Related productsstands out in the field, widely used.
Technical specifications
- Standard: SC/T 7006-2001《dissolved oxygen analyzer》;
- LCD, menu-driven operation;
- Measuring range: (0–200.0)ug/L; (0–20.00)mg/L(auto-ranging); (0–60)℃; saturation: (0–200.0)%;
- resolution: 0.1 ug/L, 0.01 mg/L, 0.1℃; 0.1%;
- complete instrument Basic error: ug/L: ±1.0%FS; mg/L: ±0.5%FS, temperature: ±0.3℃;
- complete-instrument indication Repeatability: ±0.5%FS;
- complete-instrument indication Stability: ±1.0%FS;
- Automatic temperature compensation range: (0–60)℃, referenced to 25°C;
- response time: <60 s( of full scale98%, 25℃)37℃: 98%full scale<20 s;
- Output current error: ≤±1.0%FS;
- Isolated output: (4–20)mA(loadresistance<750Ω );
- RS485communication interface(optional);
- data storage capacity: 1 months(1 pt/5 min);
- data retention after power loss: 10 years;
- alarm relay: AC220V, 3A;
- Power supply: 220V±10% 50±1HZ; optionalDC24V, please specify when ordering;
- Overall dimensions: transmitter: 159 (L)* 159 (W)* 138 (D)* mm; Cutout size: (138*138)mm
- Weight: transmitter: 0.8Kg;
- Operating conditions: Ambient temperature: 0–60℃, Relative humidity: <85%;
- inlet/outlet piping: stainless steelrigid-pipe connection; optionalφ8stainless steelquick connector.
- flowrequirements: (100–400)ml/min(6–24)L/h.
Standard configuration
| OXY5401-IXDDLtransmitter | one |
| oxygen electrode | two (XT-DOppb-DDL) |
| flow-through | two |
| instruction manual | one |
| certificate of conformity | one |
| packing list | one copy of |
Features
- fast sensor response;
- sensor measurement Stability;
- sensorlow maintenance;
- instrumentmeasurementaccurate data, Stability;
- microprocessor-based, uses high-performanceCPUchip, AccuracyADconversion technology and SMTSMT technology;
- EMC(EMC/RFI)design: designed to European StandardEN50081/50082design and manufacture;
- dustproof and waterproofdesign;
- dual-channeldesign, one instrumentmeasurement2-channel sampleoxygen content.